Operation
Selecting the ventilation mode
IPPV
42
Alarms effective in IPPV mode
– – : The default setting on delivery is outside the
monitoring range and the corresponding alarm limit
is switched off.
To set the values of the alarm limits during operation see
page 60.
To set the default alarm limits see page 86.
Alarm limits In IPPV mode Default setting
on delivery
SpO2 ON
ON
– –
92
Pulse
H
ON
ON
120
50
etCO2 ON
ON
50 (mmHg)
– –
FiCO2 ON Fixed upper alarm
limit 5 mmHg
MV ON
ON
3.0
FiO2 ON
ON
20
Fi Hal. ON
ON
1.5
– –
Fi Iso. ON
ON
2.3
– –
Fi Enf. ON
ON
3.4
– –
Fi Des. ON
ON
12.0
– –
Fi Sev. ON
ON
3.4
– –
PAW ON
ON
40
8
Apnoea pressure ON Active after
15 seconds
Apnoea flow ON Active after
15 seconds
Apnoea CO2 ON Active after
30 seconds
_
_
_
_
_
_
_
_
_
_
>
_