F9, F9 Express Fetal & Maternal Monitor User Manual Maintenance and Cleaning
- 149 -
WARNING
1 Unplug the monitor and the base station from the AC power source and detach all
accessories before cleaning. Do not immerse the unit in water or allow liquids to enter
the case.
2 If liquid is splashed on or into the main unit inadvertently, or enters the conduit, stop
using the monitor and contact the manufacturer for service immediately.
The solutions recommended for monitor cleaning are: mild near neutral detergent, ethanol 75%
and isopropanol 70%.
Clean the monitor and the base station enclosure with soft cloth and diluent non-caustic
detergents recommended above.
Clean the screen and the charging point in the docking slot with a dry soft cloth.
CAUTION
1 Although the monitor and the base station are chemically resistant to most common
hospital cleaners and non-caustic detergents, different cleaners are not
recommended and may stain the monitor.
2 Many cleansers must be diluted before use. Follow the manufacturer’s directions
carefully to avoid damaging the monitor and the base station.
3 Do not use strong solvent, for example, acetone.
4 Never use an abrasive such as steel wool or metal polish.
5 Do not allow any liquid to enter the product, and do not immerse any part of the
monitor into any liquid.
6 Avoid pouring liquids on the monitor while cleaning.
7 Do not allow any remaining solution on the surface of the monitor.
NOTE:
1 The monitor surface can be cleaned with hospital-grade ethanol and dried in air or
with crisp and clean cloth.
2 The manufacturer has no responsibility for the effectiveness of controlling infectious
disease using these chemical agents. Please contact infectious disease experts in
your hospital for details.
12.2.2 Cleaning of Accessories
(1) Cleaning of Transducers
To clean the transducers and leads, follow these steps:
1) Wipe them with a soft cloth dampened in cleaning solution;
2) Clean them with a soft cloth dampened in water;