EasyManuals Logo

EDAN F9 Express User Manual

EDAN F9 Express
215 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #165 background imageLoading...
Page #165 background image
F9, F9 Express Fetal & Maternal Monitor User Manual Product Specifications
- 154 -
Appendix 1 Product Specifications
A F9, F9 Express Fetal/Maternal Monitor
A1.1 Environmental Specifications
The monitor may not meet the performance specifications given here if stored or used outside the
specified temperature and humidity ranges.
Working
Temperature
+5 ºC ~ + 40 ºC ( +41 ºF ~ +104 ºF)
Relative Humidity
15% ~ 93% (non-condensing)
Atmospheric Pressure
86 kPa ~ 106 kPa
Transport and Storage
Temperature
-20 ºC ~ +55 ºC (-4ºF ~ +131 ºF)
Relative Humidity
15% ~ 93% (non-condensing)
Atmospheric Pressure
70 kPa ~ 106 kPa
A1.2 Physical Specifications
Monitor
Dimensions and Weight
Size (depth x width x height)
347mm x 330mm x 126mm
Weight
F9: Approx. 5.5 kg
F9 Express: Approx. 6.3 kg
Power Supply
Operating Voltage
100V-240V~
Operating Frequency
50Hz/60Hz
Input Power
1.0A-0.5A
Battery
14.8VDC/5000mAh
Standards Compliance
IEC 60601:2005+A1:2012, EN 60601-1:2006+A1:2013,
IEC 60601-1-2:2014, EN 60601-1-2:2015,
IEC/EN 60601-2-27,
IEC/EN 60601-2-37,
IEC/EN 60601-2-49,
IEC 80601-2-30,
ISO 80601-2-61,
ISO 80601-2-56,
AAMI/ANSI EC13
Anti-electric Shock Type
Class I equipment with internal power supply