EasyManuals Logo

EDAN F9 Express User Manual

EDAN F9 Express
215 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #122 background imageLoading...
Page #122 background image
F9, F9 Express Fetal & Maternal Monitor User Manual Fetal Monitoring Display
- 111 -
15 seconds.
DECELERATIONS:
the deceleration time.
SHORT TERM
VARIATION:
the short-term variation analysis result.
LONG TERM
VARIATION:
the long-term variation analysis result.
ANALYSIS START:
the start time of the analysis.
ANALYSIS END:
the finishing time of the analysis.
During 10 to 60-minute of the timer, the monitor gives CTG analysis results every minute.
At the end of the printing, the recorder prints the CTG analysis results of this moment on the
recorder paper.
Be aware that CTG analysis result is a calculation output. It can be used as a reference to assist
medical personnel in making correct diagnosis, instead of replacing it.
NOTE:
Do not disconnect the ultrasound transducer(s) before the printing stops, otherwise the
analysis results will not be printed.
8.2.6 *Marking a Note
When there is a significant event, you can press the MARK key on the front panel to add a note.
An event mark will appear on both the main interface and the recorder paper.
However, an event mark cannot clearly indicate an event. Smart Notes provides a list of
annotation for the events, including events that relate to drugs, positions, membranes, procedures,
antenatal, reasons and others.
8.2.6.1 Enabling/Disabling Smart Notes
To enable or disable Smart Notes,
1 Select the setup key on the main interface.
2 Select General > Smart Notes.
3 Select ON or OFF (default).
4 Select OK.
A smart note editing key appears next to the Smart Notes item.
8.2.6.2 Annotating an event
Once Smart Notes is enabled, press the MARK key on the front panel to open the smart note list,
choose an event catalog and then choose an annotation from the list.
The annotation of this event will be printed in the top area of recorder paper during real-time
printing.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the EDAN F9 Express and is the answer not in the manual?

EDAN F9 Express Specifications

General IconGeneral
BrandEDAN
ModelF9 Express
CategoryMedical Equipment
LanguageEnglish