EasyManua.ls Logo

EDAN F9 Express - Testing US Transducers

EDAN F9 Express
215 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
F9, F9 Express Fetal & Maternal Monitor User Manual Fetal Monitoring
- 84 -
4 Select a value from 0 ~ 300 second(s) for High Alarm Delay. (The step is 5, and the default
value is 10 seconds.)
5 Select a value from 0 ~ 300 second(s) for Low Alarm Delay. (The step is 5, and the default
value is 10 seconds.)
6 Select OK.
WARNING
The FHR alarm delay is adjustable between 0 and 300 seconds.
7.2.6 Testing US Transducers
To test a US transducer:
1 Switch on the monitor.
2 Connect the US transducer to the fetal monitor.
3 Hold the transducer with one hand, and gently touch the center of the transducer with the
other hand in the frequency of 2 times per second.
Figure 7-2 Testing a US Transducer
4 Check that the value on the display shows this change in FHR.
If a US transducer fails the test, repeat this test with another transducer. If the second one passes
the test, defect of the first transducer is confirmed. Replace it with a good one. If the second
transducer fails the test as well, contact the manufacturer for service.

Table of Contents

Related product manuals