6.5 Setting Date and Time
You can change the date and time of the monitor,
1 Select the setup key on the main interface.
2 Select Date And Time.
3 Set the year, month, date, hour, minute and second. The first three numbers are used to set
the year, month and date. Their orders vary with the preset Date Format below.
4 Select Date Format for the format of the date; there are three options: yyyy-mm-dd
(default), mm/dd/yyyy and dd/mm/yyyy.
5 Select OK.
CAUTION
1 You should set date and time information in advance. After this information is
changed, the monitor starts new monitoring with an auto ID. Therefore, we advise
you to restart the monitor after changing date or time information, and do not
perform this operation when monitoring is in process.
2 If date and time cannot be saved, it is probably the battery has reached the end of its
service life. Please contact the service personnel or your local distributor.
NOTE:
1 The date and time remain in the monitor for at least two months after it is switched
off. You do not have to set date and time each time before monitoring.
2 When the network version is ETHERNET 1.4, if the fetal monitor gets on line or time
is revised on the central station and the time difference between the fetal monitor
and the central station is bigger than 1 minute, then the fetal monitor will synchronize
time with the central station.