EasyManuals Logo

EDAN F9 Express User Manual

EDAN F9 Express
215 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #144 background imageLoading...
Page #144 background image
F9 and F9 Express Fetal & Maternal Monitor User Manual Maternal Monitoring
- 133 -
4 Select OK.
Note:
When the network version is ETHERNET 1.4,if the fetal monitor gets on line, the
central station will control to make the NIBP unit of the fetal monitor in concert with
that of the central station.
9.4.8 Switching the NIBP Alarm On or Off
You can choose to switch the NIBP alarm on or off. The SYS alarm and DIA alarm are related.
Once one of them is switched off, the rest will be switched off as well.
1 Select the setup key on the main interface.
2 Select Alarm. On the displayed Password box, enter 9999, then select Enter.
3 Select SYS, DIA or MAP.
4 Select ON (default) or OFF for Alarm.
5 Select OK.
9.4.9 Changing SYS Alarm Limits
You can change the SYS alarm limits.
1 Select the setup key on the main interface.
2 Select Alarm. On the displayed Password box, enter 9999, then select Enter.
3 Select SYS.
4 Select a value from 40 ~ 269 (mmHg) or 5.3~35.9 (kPa) for Low Alarm Limit. (If the unit is
mmHg, the step is 1, and the default value is 90 mmHg; if the unit is kPa, the step is 0.1, and
the default value is 12.0 kPa.)
5 Select a value from 41 ~ 270 (mmHg) or 5.4~36.0 (kPa)for High Alarm Limit. (If the unit is
mmHg, the step is 1, and the default value is 160 mmHg; if the unit is kPa, the step is 0.1,
and the default value is 21.3 kPa.)
6 Select OK.
9.4.10 Changing DIA Alarm Limits
You can change the DIA alarm limits.
1 Select the setup key on the main interface.
2 Select Alarm. On the displayed Password box, enter 9999, then select Enter.
3 Select DIA.
4 Select a value from 10 ~ 214 (mmHg) or 1.3~28.6 (kPa) for Low Alarm Limit. (If the
unit is mmHg, the step is 1, and the default value is 50 mmHg; if the unit is kPa, the step
is 0.1, and the default value is 6.8 kPa.)

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the EDAN F9 Express and is the answer not in the manual?

EDAN F9 Express Specifications

General IconGeneral
BrandEDAN
ModelF9 Express
CategoryMedical Equipment
LanguageEnglish

Related product manuals