F9, F9 Express Fetal & Maternal Monitor User Manual Fetal Monitoring
- 83 -
monitor will no longer give any audible or visual warning for this monitoring item.
1 Select the setup key on the main interface.
2 Select Alarm. On the displayed Password box, enter 9999, then select Enter.
3 Select FHR.
4 Select ON (default) or OFF for Alarm.
5 Select OK.
If FHR alarm is switched off, an alarm switched-off symbol will appear in the numeric
window. For example:
WARNING
Do not switch the alarm off for the condition where the patient’s safety maybe
endangered.
7.2.4 Changing the FHR Alarm Limits
You can change the FHR alarm limits. The alarm limits you set determine the conditions that
trigger the alarm.
1 Select the setup key on the main interface.
2 Select Alarm. On the displayed Password box, enter 9999, then select Enter.
3 Select FHR.
4 Select a value from 65 ~ 210 for High Alarm Limit. (The step is 5, and the default value is
160 bpm.)
5 Select a value from 60 ~ 205 for Low Alarm Limit. (The step is 5, and the default value is
110 bpm.)
6 Select OK.
7.2.5 Changing the FHR Alarm Delay
You can change the FHR alarm delay. The alarm delay indicates how long the measured result
continues exceeding its limit before the alarm is triggered.
1 Select the setup key on the main interface.
2 Select Alarm. On the displayed Password box, enter 9999, then select Enter.
3 Select FHR.