EasyManuals Logo

EDAN F9 Express User Manual

EDAN F9 Express
215 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #194 background imageLoading...
Page #194 background image
F9, F9 Express Fetal & Maternal Monitor User Manual Ultrasound Intensity and Safety
- 183 -
A4.6 Transducer Acoustic Output Parameters List
A4.6.1 Test of Wired Transducer
Acoustic output reporting table for IEC60601-2-37
(IEC60601-2-37, Edition 2.1, 2015-06, table 201.103)
Operating Mode: PW mode
Working Frequency: 1.0MHz
Index label
MI
TIS
TIB
TIC
At
surface
Below
surface
At
surface
Below
surface
Maximum index value
0.013
0.0091
0.051
N/A
Index component value
N/A
0.0091
N/A
0.051
Acoustic
Parameters
p
r.
α
at z
MI
(MPa)
0.013
P (mW)
8.22
8.22
N/A
P
1x1
(mW)
N/A
N/A
z
s
(cm)
13.10
z
b
(cm)
13.10
z
MI
(cm)
12.80
z
PII.
α
(cm)
12.80
f
awf
(MHz)
1.00
1.00
1.00
N/A
Other
Information
prr (Hz)
2000.00
srr (Hz)
N/A
n
pps
N/A
I
pa.
α
at z
PII.
α
(W/cm
2
)
0.011
I
spta.
α
at z
PII.
α
or z
SII.
α
(mW/cm
2
)
1.92
I
spta
at z
PII
or z
SII
(mW/cm
2
)
4.64
p
r.
at z
PII
(MPa)
0.020
Operating
control
conditions
Focus(mm)
Fixed
Depth(mm)
Fixed
Frequency(MHz)
1.00
Operating Mode: PW mode
Working Frequency: 1.0MHz
Acoustic Output
MI
I
SPTA.3
(mW/cm
2
)
I
SPPA.3
(W/cm
2
)
Global Maximum Value
0.013
1.92
0.011
Associated Acoustic
Parameter
P
r.3
(MPa)
0.013
W
0
(mW)
8.22
8.22
f
c
(MHz)
1.00
1.00
1.00
Z
sp
(cm)
12.80
12.80
12.80
Beam
x
-6
(cm)
0.62
0.62