EasyManuals Logo

EDAN F9 Express User Manual

EDAN F9 Express
215 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #174 background imageLoading...
Page #174 background image
F9, F9 Express Fetal & Maternal Monitor User Manual Product Specifications
- 163 -
B FTS-3 Fetal Telemetry System
B1.1 Environmental Specifications
Working
Temperature
+5 ºC ~ +40 ºC (+41ºF ~ +104ºF)
Relative Humidity
15% ~ 93% (non-condensing)
Atmospheric Pressure
86 kPa ~ 106 kPa
Transport and Storage
Temperature
-20 ºC ~ +55 ºC (-4ºF ~ +131 ºF)
Relative Humidity
15% ~ 93% (non-condensing)
Atmospheric Pressure
70 kPa ~ 106 kPa
B1.2 Physical Specifications
Power Supply
Operating Voltage
100V-240V~
Operating Frequency
50Hz/60Hz
Input Power
0.8A-0.3A
Battery
14.8VDC/5000mAh
Standards Compliance
IEC 60601:2005+A1:2012, EN 60601-1:2006+A1:2013,
IEC 60601-1-2:2014, EN 60601-1-2:2015,
IEC/EN 60601-2-37, IEC 60601-2-27, EN 62479:2010,
ETSI EN 301 489-1, ETSI EN 301 489-3, ETSI EN 300
220-1, ETSI EN 300 220-2.
Anti-electric Shock Type
Class I equipment with internal power supply
Anti-electric Shock Degree
FHR1, FHR2, TOCO BF
DECG, MHR(from MECG) CF
Degree of Protection against
Harmful Ingress of Water
Base station: IPX1 (protected against vertically falling water
drops)
Transducers: IPX8 (Protected against the effects of
continuous immersion in water 1.1m deep for 24 hours)
Degree of Safety in Presence
of Flammable Gases
Equipment not suitable for use in presence of flammable
gases
Disinfection/Sterilizing
Method
Refer to this user manual for details
EMC
CISPR11 Group 1 Class A

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the EDAN F9 Express and is the answer not in the manual?

EDAN F9 Express Specifications

General IconGeneral
BrandEDAN
ModelF9 Express
CategoryMedical Equipment
LanguageEnglish

Related product manuals