F9, F9 Express Fetal & Maternal Monitor User Manual Product Specifications
- 164 -
Leakage Current
Ground Leakage Current
(Limit)
N.C. S.F.C.
500µA 1000µA
Enclosure Leakage Current
(Limit)
N.C. S.F.C.
100µA 500µA
Patient Leakage Current
(Limit)
FHR1, FHR2, TOCO
N.C. S.F.C.
d.c. 10µA 50µA
a.c. 100µA 500µA
Patient Auxiliary Current
(Limit)
FHR1, FHR2, TOCO
N.C. S.F.C.
d.c. 10µA 50µA
a.c. 100µA 50A
Patient Leakage Current
(Limit)
DECG, MHR(from MECG)
N.C. S.F.C.
d.c. 10µA 50µA
a.c. 10µA 50µA
Patient Auxiliary Current
(Limit)
DECG, MHR(from MECG)
N.C. S.F.C.
d.c. 10µA 50µA
a.c. 10µA 50µA
Base Station
Weight
About1.8 kg
Size
310mm x 235mm x81mm
US-T Transducer
Weight
About 150 g
Size
Ø81 mm × 35 mm
TOCO-T Transducer & TOCO-E Transducer
Weight
About 150 g
Size
Ø81 mm × 35 mm