F9, F9 Express Fetal & Maternal Monitor User Manual Product Specifications
- 159 -
Range: 7s ~ 8s (average: 7.5s)
MHR range: 80bpm ~ 40bpm
Range : 7s ~ 8s (average: 7.5s)
Exceeds ANSI/AAMI EC13-2002 Sect. 3.1.2.1
(C)
minimum recommended 1.2mV T-Wave
amplitude
*Measuring Accuracy
(EDAN)
90% ~ 100% ± 2%
70% ~ 90% ± 4%
< 70% unspecified
*Measuring Accuracy
(Nellcor)
70% ~ 100% ± 2%
< 70% unspecified
*Data update period
(EDAN)
*Data update period
(Nellcor)
Infrared light: (90510) nm
Emitted light energy: < 15 mW
Information about the wave length range can be especially useful to clinicians
(for instance, when photodynamic therapy is performed.)
Systolic Pressure, Diastolic Pressure, Mean Artery
Pressure
Systolic Pressure: 40 mmHg ~ 270 mmHg (5.3
kPa~36.0 kPa)
Diastolic Pressure: 10 mmHg ~ 215 mmHg (1.3