F9, F9 Express Fetal & Maternal Monitor User Manual Alarms
- 62 -
**TEMP LOW or
**TEMP xxx < yyy
TEMP
TEMP result (xxx) is lower than
the lower limit (yyy).
Technical Alarm Messages
Alarm Message
Source
Cause
Countermeasure
High Level
***ECG SINGNAL
EXCEEDS LIMIT
ECG
ECG signal exceeds the
measurement limits.
Check the connection of the
leads and the patient’s
condition.
Low Level
Signals Overlap
(FHR1, HR)
US+ECG
/Pulse
US transducer 1 has picked up
the maternal heart signal; the
signals overlap.
Reposition the US transducer
1 until the fetal heart signal is
detected.
Signals Overlap
(FHR2, HR)
US+ECG
/Pulse
US transducer 2 has picked up
the maternal heart signal; the
signals overlap.
Reposition the US transducer
2 until the fetal heart signal is
detected.
Signals Overlap
(FHR1, FHR2,
HR)
US+ECG
/Pulse
US transducer 1 and US
transducer 2 have picked up the
maternal heart signal; the signals
overlap.
Reposition the US
transducers until the fetal
heart signals are detected.
ECG LEADS OFF
ECG
The ECG cable is not well
connected to the monitor or the
TOCO-E transducer, or the leads
of the ECG cable are not well
attached to the patient.
Check the connection
between the ECG cable and
the monitor or the TOCO-E
transducer. Check also the
attachment of the ECG leads
to the patient.
ECG EQUIP
MALF
ECG
The ECG board cannot
communicate with the system
successfully.
Restart the monitor and try
again, contact the
manufacturer if the
connection still fails.
HR EXCEEDS
MEASUREMENT
RANGE
ECG/
Pulse
The heart rate exceeds the
measurement limits.
Check the connection of the
ECG leads/SpO
2
transducer
and the patient’s condition.
NIBP EQUIP
MALF
NIBP
The NIBP board cannot
communicate with the system
successfully.
Restart the monitor and try
again, contact the
manufacturer if the
connection still fails.
NIBP SYSTEM
FAILURE
NIBP
The NIBP module defective.
Restart the monitor and try
again, contact the
manufacturer if the
connection still fails.
NIBP CUFF
LOOSE or OFF
NIBP
The cuff is loose or not
connected.
Wrap the cuff properly.