Promag 53 PROFIBUS DP/PA Troubleshooting
Endress+Hauser 139
572 P: HOLD BATCHING
!: # 572
0x80 GOOD O.K. O.K. Batch hold Cause:
Batching has been interrupted (valves are
closed)
Remedy:
Using either:
•PROFIBUS DP/PA
•Local display:
– Continue batching with "GO ON".
– Interrupt batching with "STOP".
No. # 6xx Simulation mode active
601 S: POS. ZERO-RETURN
! # 601
0x53 UNCERTAIN Sensor
conversion
not accurate
(measured
value from
sensor not
accurate)
Constant Positive zero
return active
Cause:
Positive zero return is active.
!
Note!
This message has the highest display priority!
Remedy:
Switch off positive zero return.
Access:
BASIC FUNCTION SYSTEM PARAMETER
CONFIGURATION POS. ZERO RETURN (
OFF)
611
to
614
S: SIM. CURR. OUT. n
!: # 611 to 614
0x80 GOOD O.K. O.K. Simulation IO
active
Cause:
Simulation current output is active.
Remedy:
Switch off simulation.
621
to
624
S: SIM. FREQ. OUT. n
!: # 621 to 624
0x80 GOOD O.K. O.K. Simulation IO
active
Cause:
Simulation frequency output is active.
Remedy:
Switch off simulation.
631
to
634
S: SIM. PULSE n
!: # 631 to 634
0x80 GOOD O.K. O.K. Simulation IO
active
Cause:
Simulation pulse output is active.
Remedy:
Switch off simulation.
641
to
644
S: SIM. STAT. OUT n
!: # 641 to 644
0x80 GOOD O.K. O.K. Simulation IO
active
Cause:
Simulation status output is active.
Remedy:
Switch off simulation.
651
to
654
S: SIM. RELAY n
!: # 651 to 654
0x80 GOOD O.K. O.K. Simulation IO
active
Cause:
Simulation relay output is active.
Remedy:
Switch off simulation.
671
to
674
S: SIM. STATUS IN n
!: # 671 to 674
0x80 GOOD O.K. O.K. Simulation IO
active
Cause:
Simulation status output is active.
Remedy:
Switch off simulation.
No. Device status message
(local display)
PROFIBUS measured value status
Extended
diagnostic
message in the
PROFIBUS
master
Cause/remedy
(Spare parts → 146)
Quality code (HEX)
Measured value status
Quality status
Quality substatus
Limits