8.  USB Computer Port Connector, Type B (J16):  This connector is available to 
interface the H201Ci-1 directly to a PC or to a smart modem.  The RS-232 
computer port (J10) can also be used (item 7 in Figure 4-7. 
 
9.  5-Vdc and 12-Vdc Power Supply Connector, 4 Pins (J6):  This power source can be 
used for optional equipment. Each line is protected with a 1.35A fuse. 
 
10.  Optional  Modem Connector, 4 Pins  (J9):   This connector  is used  to interface the 
H201Ci-1 with a modem.  The telephone line should be connected here.  For more 
detail, refer to the modem procedure located in the English/Procedures folder of 
the Hydran 201i System installation CD. 
 
11.  RS-485 LAN Connector, 3 Pins (J11):  To set up an H201Ti network, the H201Ci-1s 
(and/or  H201Ci-Cs)  are  daisy-chained  together  into  an  RS-485  Local  Area 
Network (LAN). 
 
12.  Supervisory  Link  Connector,  6  Pins  (J15):    The  isolated  supervisory  link  cable 
coming from the H201Ti is connected here.  The link cable carries the TDM pulse 
signal (analog data and relay status), serial data (communications) and isolated 
loop supply (+15 Vdc, from the H201Ci-1). 
 
13.  Heater Indicator, Yellow (DS8):  This LED indicator is lit when the enclosure heater 
is on.  The enclosure temperature is controlled and the heater periodically turns 
on and off to maintain the temperature near 25°C (77°F). 
 
14.  Alarm  2 (Gas High-High) Indicator,  Yellow (DS9):  This LED indicator is lit when a 
Gas High-High alarm condition (level, hourly trend or daily trend) is detected by 
the H201Ti.  It turns off when the condition is cleared.  The H201Ci-1’s Gas High-
High relay tracks the equivalent relay in the H201Ti. 
 
15.  Alarm 1 (Gas High) Indicator, Yellow (DS10):  This LED indicator is lit when a Gas 
High alarm condition (level, hourly trend or daily trend) is detected by the H201Ti.  
It turns off when the condition is cleared.  The H201Ci-1’s Gas High relay tracks 
the equivalent relay in the H201Ti. 
 
16.  System  OK  Indicator,  Yellow  (DS11):    This  LED  indicator  is  lit  when  there  is  no 
system fault.  It turns off when a system fault is detected by the H201Ti, or when 
the link is broken between the H201Ti and the H201Ci-1, or if there is an ac. power 
loss  in  the  H201Ti  or  the  H201Ci-1.    The  H201Ci-1’s  System  Fail  relay  is  de-
energized when the System OK indicator is off. 
 
17.  Supervisory Link Supply Indicator, Green (DS12):  This LED indicator is lit when the 
supervisory link power supply is On. 
 
18.  Analog Out Failure Indicator, Red-Orange (DS13):  This LED indicator is lit when the 
analog output is not working properly, is unplugged or is not used.