5. Do not paint the sensor (see Figure 6-3). Paint fumes block the sensor’s breathing
membrane and damage the sensor permanently, thus voiding the warranty.
6. Do not use hydrocarbon-based compounds near the sensor. Do not clean the
sensor with any solvent or other products (see Figure 6-3 below). The volatile
fumes of these compounds can deteriorate sensor performances. Products to
avoid include: Paint, liquid Teflon, vehicle exhaust, spray can, black pitch, thinner,
RTV (Room Temperature Vulcanization; silicon-based mastic) and solvent.
Figure 6-3: Do not paint the Sensor or do not clean it with any solvent