Chapter 8: System Hardware Overview
5495975-1EN Rev.9 8-17
© 2013-2017 General Electric Company. All rights reserved.
17. Control All
Transitional
Locks
Released
LED
This green light is lit up when all lock release is active except for
tube angulation.
18. Control Screen OTS Control
Screen
Displays and controls exposure settings.
Refer to OTS Control Screen
for more information.
19. Control Key Switch Located on the back of the OTS User Interface, this key switch is
intended for use when a system failure has occurred. The
OVERRIDE position disables the exposure hold interlocks and
allows exposure. In the digital wall stand applications, the full
detector area and SID = 100cm is used when in OVERRIDE position.
Engaging OVERRIDE mode shuts down certain system functions,
including Auto-tracking and Auto-collimation. This may
effect image quality.
If OVERRIDE is engaged, return the switch to the NORMAL position
at the end of the exam.
1) Normal Position 2) OVERRIDE Position
CAUTION If the key switch OVERRIDE is selected, you
must ensure the collimator field of view
matches the receptor and the tube is aligned
to the selected receptor. Use the collimator
field light to match the X-ray field to the
receptor.
Table 8-6 OTS Function
Item Number and
Type
Description Description