JC-TSG-A-007.fm
GE Healthcare Senographe DS
Revision 1 Service Information and Procedures Class A 2385072-16-8EN
Job Card TSG A007 - Digital Detector Firmware Corrupted
execute the loadDetectorFW command you typed in step 4.
The flashing/"firmware write" occurs during this minute.
6. Some messages on the screen indicating that the IDC must be rebooted.
7. At the -> prompt, reboot the IDC by entering:
idcReboot
The detector should boot normally.