GE Healthcare Senographe DS
Revision 1 Service Information and Procedures Class A 2385072-16-8EN
Job Card D/R A315 - International Keyboard
Page no. 1443 Chapter 9
JC-DR-A-315.fm
Job Card D/R A315 - International Keyboard Chapter 9
1 APPLICABILITY
This Job Card is only applicable to V3 and V4 Control Stations.
Perform this Job Card if you want to:
• change the US Keyboard that is delivered with the Senographe system to an International Keyboard
(i.e. keyboard with a different layout to the US layout)
• change an existing keyboard (either US Keyboard or International Keyboard) for a keyboard with the
same layout.
Note:
You must not use a Sun keyboard with a V4 Control Station (Z400), since it can lead to keyboard
malfunction.
2SUPPLIES
For V3 Control Stations, one of the following keyboards:
For V4 Control Stations, one of the following keyboards:
Keyboard Layout FRU (GE Part Number)
German 5233716-2
Swedish 5233716-3
Norwegian 5233716-4
Spanish 5233716-5
French 5233716-6
Portuguese 5233716-7
Italian 5233716-8
Danish 5233716-9
Dutch 5233716-10
Finnish 5233716-11
US (default) 5233716
Keyboard Layout FRU (GE Part Number)
German 5394524-2
Swedish / Finnish 5394524-3
Norwegian 5394524-4
Spanish 5394524-5
French 5394524-6
Portuguese 5394524-7
Italian 5394524-8
Danish 5394524-9
Dutch 5394524-10
US (default) 5394524

Table of Contents

Other manuals for GE Senographe DS

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Senographe DS and is the answer not in the manual?

GE Senographe DS Specifications

General IconGeneral
BrandGE
ModelSenographe DS
CategoryMedical Equipment
LanguageEnglish

Related product manuals