GE Healthcare Senographe DS
Revision 1 Service Information and Procedures Class A 2385072-16-8EN
Job Card D/R A203 - Lower Optical Fork
Page no. 1171 Chapter 9
JC-DR-A-203.fm
Job Card D/R A203 - Lower Optical Fork Chapter 9
1SUPPLIES
Lower Optical Fork
2TOOLS
Standard Tool Box
3 REQUIRED EFFORT
Personnel: 1 Field Engineer
Time: 25 minutes
4 SAFETY PRECAUTIONS
WARNING
The Gantry must be powered off.
5 PREREQUISITES
Raise the arm to its maximum height to access the lower
optical fork (1).
Switch off the system power.
Remove the Front cover and the right column cover (Refer
to Job Card PHY A044 - Remove/Reinstall Gantry Covers
on page 523).
Loosen and remove the Gantry curtain under the arm (refer
to Job Card D/R A208 - Gantry Curtains on page 1193).