Job Card PM A155 - General System Inspection Checks
Page no. 1775 Chapter 11
JC-PM-A-155.fm
Job Card PM A155 - General System Inspection Checks Chapter 11
1SUPPLIES
None
2TOOLS
None
3 REQUIRED EFFORT
Personnel: 1 Field Engineer
Time: 10 minutes
4 SAFETY PRECAUTIONS
None
5 PREREQUISITES
None
6 PROCEDURE
6-1 Other General Inspection Checks
6-1-1 Visual inspection of cables
Check the appearance of all visible cables and correct hazardous trailing cables in the room.
6-1-2 Visual inspection of equipment
Note any signs of damage to covers, monitor support, etc.
6-1-3 Inspect the Radiation Screen
Check for breaks or cracks in the glass, gaps or cracks in the frame, and for correct, stable mounting,
etc.
6-2 System Cover Checks
6-2-1 Generator cover check
Check covers for security and presence of screw caps, and for damage, scratches, etc.
6-2-2 Gantry cover check
Check covers for security and presence of screw caps, and for damage, scratches, etc.
6-2-3 Control Station cover check
Check covers for security and presence of screw caps, and for damage, scratches, etc.