GE Healthcare Senographe DS
Revision 1 Service Information and Procedures Class A 2385072-16-8EN
Job Card CHK A011 - Check kVp settings
Page no. 1745 Chapter 9
JC-CHK-A-011.fm
Job Card CHK A011 - Check kVp settings Chapter 9
1SUPPLIES
None
2TOOLS
kVp meter. The kVp meter must be in good condition and possess valid certification.
Suitable meters are:
- Radcal 4083 kVp meter
- RMI 232 kVp meter
- Keithley 35080B kVp meter
Note:
The calibration curve attached to the meter (obtained during periodic recalibration of the meter)
must be used to correct the meter reading throughout this procedure.
Note:
The Keithley 35080B kVp meter is not recommended for use in Europe
It does not meet the European accuracy requirement of 3%.
Standard 19 X 23 Compression Paddle for positioning the kVp meter close to the Tube Head
Detector Protection Plate (used to protect the detector from the possibility of falling objects)
Radiation Protection Plate (used to protect the detector from excessive exposure to X-rays)
3 REQUIRED EFFORT
Personnel: 1 Field Engineer
Time: 5 minutes
4 SAFETY PRECAUTIONS
The procedures in this Job Card produce X-rays. Ensure that you take appropriate precautions when
operating with X-rays (for example, stay behind the radiation screen during an X-ray exposure).
5 PREREQUISITES
None
6 PROCEDURE
The procedures described below are designed to verify that the values of kVp delivered by the equip-
ment correspond to the values selected by the operator.