GE Healthcare Senographe DS
Revision 1 Service Information and Procedures Class A 2385072-16-8EN
Job Card ELE A028 - Collimation Checks
Page no. 679 Chapter 7
JC-ELE-A-028.fm
Job Card ELE A028 - Collimation Checks Chapter 7
1 OBJECTIVE
To ensure that:
there is no excessive extension of the X-ray field beyond the edges of the image receptor
the X-ray field aligns with the light field,
the chest wall edge of the compression paddle aligns with the chest wall edge of the image receptor
2SUPPLIES
None
3TOOLS
One coin (e.g., a nickel, or a 5-Eurocent coin).
4 strips of GAFCHROMIC® XR-M film for each FOV and track tested
Aluminium attenuator plate
Radiation Protection Plate (used to protect the detector from excessive exposure to X-rays)
4 REQUIRED EFFORT
Personnel: 1 Field Engineer
Time: 20 minutes
5 SAFETY PRECAUTIONS
The procedures in this Job Card produce X-rays. Ensure that you take appropriate precautions when
operating with X-rays (for example, stay behind the radiation screen during an X-ray exposure).
6 PREREQUISITES
None