GE Healthcare Senographe DS
Revision 1 Service Information and Procedures Class A 2385072-16-8EN
Pre-Installation System Requirements
Page no. 359 Chapter 4
Chap-Pre-Requirements.fm
11. INSITE CONNECTION
A broadband Internet connection or a Dedicated Service Network must be provided for access to Insite
services.
There are currently three main methods of providing this connection. Either Virtual Private Network
(VPN) tunneling on a broadband Internet connection, or a Dedicated Service Network as follows:
• Site to Site VPN (GE Solution) : using an analog or digital broadband router supplied by GE.
• Site to Site VPN (Customer Solution) : using an analog or digital broadband router and infrastructure
supplied by the customer.
• Dedicated Service Network provided by the country local health service (if available) (e.g. NHSnet/
N2/N3 in the UK, SJUnet in Sweden, and Sescam in Spain).
More information can be found in the Insite Pre-Installation Manual 2327624-100 and the Broad Band
(BB) Solutions Catalogue. You can download this from the OTR & Sales section at http://supportcen-
tral.ge.com/products/sup_products.asp?prod_id=24026 (GEMS SSO login credentials required).
Note:
The Insite Pre-Installation Manual and the BB Solutions Catalogue are only available via the
GEMS Intranet. If a customer require these documents then GEMS personnel can provide the cus-
tomer with these documents at request. In the Americas region, only the Site to Site VPN (Cus-
tomer Solution) is possible.
12. NETWORKING CONNECTIONS
• The Control Station and any optional equipment provided are to be connected together as a hospital
network.
• Before installation, the following information must be obtained for each network host so that it can be
addressed by the AWS:
- IP address; Gateway address; Subnet mask.
The hospital network administrator usually supplies this information.
• Provision must be made for Ethernet cables to be easily run from the Control Station to the hosiptal
network.
Typical equipment options which can be connected to the hosptal network include: review workstation,
Mass Archiver, Laser Printer, HIS/RIS network kit, CAD (Computer Aided Detection).
13. TELEPHONE CONNECTION
It is recommended that a telephone is provided close to the X-ray Console (normally mounted on the
Control Station), to allow convenient dialog with teleservice technicians.
14. ASSOCIATED REVIEW WORKSTATIONS
Planning information for the Seno Advantage Review Workstation can be found in the Volume Share/
AW/SA Pre-Installation Manual, 5180573-x-100.

Table of Contents

Other manuals for GE Senographe DS

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Senographe DS and is the answer not in the manual?

GE Senographe DS Specifications

General IconGeneral
BrandGE
ModelSenographe DS
CategoryMedical Equipment
LanguageEnglish

Related product manuals