EasyManuals Logo

GE Senographe DS User Manual

GE Senographe DS
1810 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #1017 background imageLoading...
Page #1017 background image
GE Healthcare Senographe DS
Revision 1 Service Information and Procedures Class A 2385072-16-8EN
Job Card TSG A009 - AWS Bootup Troubleshooting
Page no. 1017 Chapter 9
JC-TSG-A-009.fm
Job Card TSG A009 - AWS Bootup Troubleshooting Chapter 9
1 INTRODUCTION
This Job Card is intended as an aid to troubleshooting Acquisition Workstation (AWS) bootup problems.
It is used when a bootup fault prevents the AWS booting up correctly.
1-1 Blank Screen after Logging in as the sdc User
Once someone has logged in at the first AWS Login: prompt with the sdc user, the monitor goes blank
(white with no text or graphics) and then returns to the Login: prompt. This scenario occurs in the follow-
ing situations:
• the mouse or trackball is disconnected
• the mouse or trackball is faulty
If you witness a blank screen that keeps returning to the Login: prompt after logging in as the sdc user
then you must do the following:
• Check that the mouse/trackball is correctly attached to the AWS. Check both the connections on the
back panel of the ADS computer and on the USB extension cable that connects the mouse/trackball
to the ADS computer.
• If the problem persists after you have check the cable connections, the mouse/trackball is faulty. You
must change the mouse/trackball according to Job Card D/R A312 - Mouse on page 1417 or Job
Card D/R A313 - Trackball on page 1427.

Table of Contents

Other manuals for GE Senographe DS

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Senographe DS and is the answer not in the manual?

GE Senographe DS Specifications

General IconGeneral
BrandGE
ModelSenographe DS
CategoryMedical Equipment
LanguageEnglish

Related product manuals