GE Healthcare Senographe DS
Revision 1 Service Information and Procedures Class A 2385072-16-8EN
Job Card D/R A248 - Force Sensor
Page no. 1313 Chapter 9
JC-DR-A-248.fm
Job Card D/R A248 - Force Sensor Chapter 9
1SUPPLIES
Force Sensor
2TOOLS
Standard Tool Box
3 REQUIRED EFFORT
Personnel: 1 Field Engineer
Time: 60 minutes
4 SAFETY PRECAUTIONS
WARNING
The Gantry must be powered off.
5 PREREQUISITES
Switch off the Gantry power.
Remove the Paddle.
Remove the Paddle Holder cover (refer to Job Card
PHY A044 - Remove/Reinstall Gantry Covers on page
523) for access to the Force Sensor (1)
1