Chapter 4 Page no. 324
Chap-Pre-Requirements.fm
GE Healthcare Senographe DS
Revision 1 Service Information and Procedures Class A 2385072-16-8EN
Pre-Installation System Requirements
1. ENVIRONMENTAL REQUIREMENTS
1-1. Atmospheric Pressure Limits (Storage and Operational)
* - During transportation a pressurized environment must be used to maintain the atmospheric
pressure limits.
1-2. Preventing Thermal Shock to the Detector
To prevent thermal shock to the detector, the ambient tempertaure gradient must not exceed the values
as specified below.
1-3. Storage Requirements - temperature and humidity
1-3-1. General requirements
- The system includes a detector assembly in its casing, which is sensitive to changes in tempera-
ture and humidity, and is water-cooled when in use.
- The specified storage requirements assume that the all the equipment remains in its packaging,
including the protection for the detector.
1-3-2. Before installation - short term
For short-term storage (less than 5 days), refer to the storage requirements table below.
1-3-3. Before installation - long term
For long-term storage (more than 5 days), it is recommended that the detector assembly is kept in an
area with relatively dry and low temperature, or area with air conditioning. For example, relative humidity
less than 50%.
Atmospheric pressure
(operation, storage and transportation*)
Altitude (from sea level)
Min. Max. Min. Max.
700 hPa 1060 hPa 0 m 3000 m
0 ft 9840 ft
Temperature range Less than 0°C 0°C to 40°C Greater than 40°C
Less than 32°F 32° to 104°F Greater than 104°F
Ambient temperature
gradient must not
exceed
4.5°C per hour 1.5°C per minute 4.5°C per hour
40.1°F per hour 34.7°F per minute 40.1°F per hour
Relative humidity
(non-condensing)
Temperature
Min. Max. Min. Max.
5% 95% –20°C 50°C
–4ºF 122ºF