GE Healthcare Senographe DS
Revision 1 Service Information and Procedures Class A 2385072-16-8EN
Pre-Installation System Requirements
Page no. 325 Chapter 4
Chap-Pre-Requirements.fm
1-4. Operating Requirements - temperature and humidity
1-4-1. Air conditioning
- Air conditioning must be provided where necessary to ensure that no part of the equipment
(including the generator cabinet) operates in an ambient temperature exceeding 35°C (95ºF).
- For patient comfort, ambient temperatures of 23°C ± 3°C (73ºF ± 5ºF) are recommended.
1-5. Short Term Shutdown Requirements - temperature and humidity
1-5-1. Shutdown with Conditioner on
During evenings and weekends some Operators shutdown the Senographe system via the Control Sta-
tion; in that case the Conditioner remains switched on, and the environment must be within the following
limits.
1-5-2. Shutdown with Conditioner off
During evenings and weekends some Operators shutdown the Senographe system via the circuit
breaker on the wall; in that case the Conditioner is switched off, and the environment must be within the
following limits.
1-5-3. Air conditioning
Air conditioning must be provided where necessary to ensure that no part of the equipment (including the
generator cabinet) exists in an ambient temperature exceeding 35°C (95ºF) or below 15°C (59ºF).
Relative humidity (non-condensing) Temperature
Min. Max. Min. Max.
10% 80% 15°C 35°C
59ºF 95ºF
Relative humidity (non-condensing) Temperature
Min. Max. Min. Max.
10% 80% 15°C 35°C
59ºF 95ºF
Relative humidity (non-condensing) Temperature
Min. Max. Min. Max.
5% 95% 10°C 50°C
50ºF 122ºF