GE Healthcare Senographe DS
Revision 1 Service Information and Procedures Class A 2385072-16-8EN
Job Card PHY A012 - X-ray Console Cable Routing
Page no. 451 Chapter 6
JC-PHY-A-012.fm
Job Card PHY A012 - X-ray Console Cable Routing Chapter 6
1SUPPLIES
None
2TOOLS
Standard Tool Box
3 SAFETY PRECAUTIONS
None
4 REQUIRED EFFORT
Personnel: 1 Field Engineer
Time: 30 minutes
5 PREREQUISITES
None
Notice:
- To avoid unwanted disconnection, the X-ray Console cable must be fixed to the mechanical struc-
ture of the Control Station and the Generator.
- The external part of the X-ray Console cable must be mechanically protected from damage that
could come from unsafe routing in the room.