Chapter 6 Page no. 450
JC-PHY-A-011.fm
GE Healthcare Senographe DS
Revision 1 Service Information and Procedures Class A 2385072-16-8EN
Job Card PHY A011 - Trackball Option Installation
21. Turn the rotative arm though all extremes, and check that the harness, monitor cables, and trackball
cable are free to move as indicated below.
22. Reinstall the Omega Cover (see Job Card PHY A040 - Remove/Reinstall Control Station Covers on
page 503).
23. Reinstall the Top Cover (see Job Card PHY A040 - Remove/Reinstall Control Station Covers on
page 503).
24. Reinstall the X-ray Console.