Revision 1 Service Information and Procedures Class A 2385072-16-8EN
Job Card D/R A208 - Gantry Curtains
Page no. 1193 Chapter 9
JC-DR-A-208.fm
Job Card D/R A208 - Gantry Curtains Chapter 9
1SUPPLIES
Kit containing upper and lower Gantry Curtains
2TOOLS
Standard Tool Box
3 REQUIRED EFFORT
Personnel: 1 Field Engineer
Time: 60 minutes
4 SAFETY PRECAUTIONS
WARNING
While performing this procedure, you must:
Power off the Gantry.
Wear safety glasses and gloves.
5 PREREQUISITES
Switch off the Gantry power.
Remove the front and column covers (refer to Job Card PHY A044 -
Remove/Reinstall Gantry Covers on page 523.
Switch on the Gantry power.
Use the lift command buttons to move the arm vertically until the two
springs (1) are fully visible in the opening at the rear of the column
frame.
Switch off the Gantry power.
1