Job Card TSG A010 - Troubleshooting Bad Pixel Calibration Failure
Page no. 1027 Chapter 9
JC-TSG-A-010.fm
4-3 Calibration to be performed after an X-Ray tube mA measurement
4-4 Calibration to be performed after repositioning the X-Ray tube housing
Calibration/Test DS and Essential Job Card
X-Ray Tube mA Measurement
Job Card CAL A009 - X-Ray Tube mA Measurement on page 1663
X-Ray Tube Heater current Calibration Job Card CAL A012 - X-Ray Tube Heater Current on page 1673
Bad Pixel Calibration Job Card CAL A041 - Bad Pixel Calibration on page 1693
Gain calibration Job Card CAL A043 - Detector Gain Calibration on page 1697
Flat field tests Job Card ELE A032 – IQ Tools Flat Field Test on page 699
Conversion Factor Job Card CAL A045 - Conversion Factor Measurement on page 1701
Half Value Layer Job Card CAL A046 - Half-Value Layer Measurement on page 1703
AOP Calibration Job Card CAL A051 - AOP Calibration on page 1723
mAs non linearity Job Card CAL A053 - mAs Non-linearity Calibration on page 1725
AOP/SNR Check Job Card ELE A038 - AOP Mode and SNR Check on page 707
If AOP/SNR test is not OK, repeat all calibrations from AOP Calibration
Calibration/Test Job Card
Tube Tilt Calibration Job Card CAL A027 - Tube Tilt Calibration on page 1681
Collimator Format Calibration Job Card CAL A029 - Collimator Format Calibration on page 1683
Bad Pixel Calibration Job Card CAL A041 - Bad Pixel Calibration on page 1693
Gain calibration Job Card CAL A043 - Detector Gain Calibration on page 1697
Flat field tests Job Card ELE A032 – IQ Tools Flat Field Test on page 699
Conversion Factor Job Card CAL A045 - Conversion Factor Measurement on page 1701
Half Value Layer Job Card CAL A046 - Half-Value Layer Measurement on page 1703
AOP Calibration Job Card CAL A051 - AOP Calibration on page 1723
mAs non linearity Job Card CAL A053 - mAs Non-linearity Calibration on page 1725
AOP/SNR Check Job Card ELE A038 - AOP Mode and SNR Check on page 707
If AOP/SNR test is not OK, repeat all calibrations from AOP Calibration

Table of Contents

Other manuals for GE Senographe DS

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Senographe DS and is the answer not in the manual?

GE Senographe DS Specifications

General IconGeneral
BrandGE
ModelSenographe DS
CategoryMedical Equipment
LanguageEnglish

Related product manuals