EasyManua.ls Logo

GE Senographe DS

GE Senographe DS
1810 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
GE Healthcare Senographe DS
Revision 1 Service Information and Procedures Class A 2385072-16-8EN
Job Card TSG A016 - Rotation Movement Troubleshooting
Page no. 1065 Chapter 9
JC-TSG-A-016.fm
3-4-3 Locate dry surfaces
Rotate the examination arm between -165°
and +185°.
Check for dry or partly dry surfaces as shown
here (4); this is the visible effect of incorrect
greasing
!
Notice:
Be sure to identify dry or partly dry surfaces before greasing. General greasing may not help! The
most critical areas are visible with the arm in the following positions: 0°, 45°, 135°, and 180°.
4

Table of Contents

Other manuals for GE Senographe DS

Related product manuals