Revision 1 Service Information and Procedures Class A 2385072-16-8EN
Job Card TSG A016 - Rotation Movement Troubleshooting
3-7 Rotation Motor Checks
Check cabling connections and connectors on the Rotation Board, and the Rotation Motor.
• If the Rotation Motor connectors are not OK, change the Rotation Motor according to Job Card D/R
A228 - Rotation Motor on page 1243.
Note:
To confirm whether the Rotation Motor is faulty you can additionally measure the resistance be-
tween the motor connector pins (J10 on the Rotation Board). If the resistance is infinite (open cir-
cuit), the Rotation Motor is faulty.
• If the Rotation Board connectors are not OK, change the Rotation Board according to Job Card D/R
A221 - Rotation Board on page 1223.
• If the cabling connections and connectors are OK but the Rotation does not still work, change the
Rotation Board according to Job Card D/R A221 - Rotation Board on page 1223, since it is the last
remaining part that can be replaced.

Table of Contents

Other manuals for GE Senographe DS

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Senographe DS and is the answer not in the manual?

GE Senographe DS Specifications

General IconGeneral
BrandGE
ModelSenographe DS
CategoryMedical Equipment
LanguageEnglish

Related product manuals