EasyManua.ls Logo

GE Senographe DS

GE Senographe DS
1810 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Chapter 9 Page no. 1068
JC-TSG-A-016.fm
GE Healthcare Senographe DS
Revision 1 Service Information and Procedures Class A 2385072-16-8EN
Job Card TSG A016 - Rotation Movement Troubleshooting
3-7 Rotation Motor Checks
Check cabling connections and connectors on the Rotation Board, and the Rotation Motor.
If the Rotation Motor connectors are not OK, change the Rotation Motor according to Job Card D/R
A228 - Rotation Motor on page 1243.
Note:
To confirm whether the Rotation Motor is faulty you can additionally measure the resistance be-
tween the motor connector pins (J10 on the Rotation Board). If the resistance is infinite (open cir-
cuit), the Rotation Motor is faulty.
If the Rotation Board connectors are not OK, change the Rotation Board according to Job Card D/R
A221 - Rotation Board on page 1223.
If the cabling connections and connectors are OK but the Rotation does not still work, change the
Rotation Board according to Job Card D/R A221 - Rotation Board on page 1223, since it is the last
remaining part that can be replaced.

Table of Contents

Other manuals for GE Senographe DS

Related product manuals