Revision 1 Service Information and Procedures Class A 2385072-16-8EN
Job Card TSG M028 - E24 084/008 Error Troubleshooting
Page no. 1091 Chapter 9
JC-TSG-M-028.fm
7-3 Supported Pre-Exposure Configurations
The AOP pre-exposure mAs value is equal to 4mAs.
The AOP pre-exposure focal spot triplets are as follows:
Large focal spot Small focal spot
Track kV Track kV
Mo 22 Mo 22
Mo 23 Mo 23
Mo 24 Mo 24
Mo 25 Mo 27
Rh 27 Mo 27
Rh 28 Mo 30
Rh 31 Rh 29
Rh 32 Rh 31
Rh 34 Rh 34