Chapter 9 Page no. 1098
JC-DR-A-121.fm
GE Healthcare Senographe DS
Revision 1 Service Information and Procedures Class A 2385072-16-8EN
Job Card D/R A121 - Optical Fiber (Gantry to Generator)
10. Replace this fiber, respecting connector color coding.
11. Temporarily reconnect the four fibers to the generator and Gantry, and check that the system oper-
ates correctly when the electrical power is switched on.
12. Disconnect the four fibers and reinstall the ribbed outer cover in the harness.
13. Carefully install new cable ties, particularly inside the Gantry and between the base plate and Gantry
interface board PL104.
6-2 Functional Check
1. Switch on the Senographe electrical power.
2. Raise and lower the column and visually check that the movement is not fouling any of the cables.
3. Reinstall the generator covers and the Gantry column covers.
7 COMPLETION