EasyManuals Logo

GE Senographe DS User Manual

GE Senographe DS
1810 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #1237 background imageLoading...
Page #1237 background image
GE Healthcare Senographe DS
Revision 1 Service Information and Procedures Class A 2385072-16-8EN
Job Card D/R A224 - Gas Spring for Rotation Worm Gear
Page no. 1237 Chapter 9
JC-DR-A-224.fm
3. By hand, carefully pull the old Gas Spring (5) to remove it from the mounting plate (3).
4. By hand, carefully push the new Gas Spring (6) to into the mounting plate (3).
3
5
3
6
6
3

Table of Contents

Other manuals for GE Senographe DS

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Senographe DS and is the answer not in the manual?

GE Senographe DS Specifications

General IconGeneral
BrandGE
ModelSenographe DS
CategoryMedical Equipment
LanguageEnglish

Related product manuals