Chapter 9 Page no. 1280
JC-DR-A-233.fm
GE Healthcare Senographe DS
Revision 1 Service Information and Procedures Class A 2385072-16-8EN
Job Card D/R A233 - Lift Screw Assembly Top Brake
7 PROCEDURE
7-1 Disassembly of the old Lift Screw Assembly Top Brake
1. Remove the right hand side Gantry Column Covers to reveal the location of the Lift Screw Assembly
Top Brake and the Lift Board (refer to Job Card PHY A044 - Remove/Reinstall Gantry Covers on
page 523).
2. Disconnect the Lift screw Assembly Top Brake cable (W510) from the J6 connector on the Lift Board.
Cut the cable ties to secure the Lift screw Assembly Top Brake cable (W510) up the side of the Gan-
J6
To Brake

Table of Contents