EasyManua.ls Logo

GE Senographe DS

GE Senographe DS
1810 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Chapter 9 Page no. 1504
JC-DR-A-321_DS.fm
GE Healthcare Senographe DS
Revision 1 Service Information and Procedures Class A 2385072-16-8EN
Job Card D/R A321 - IDC Computer
6 PROCEDURE
6-1 Summary
The following list summarizes the steps described in this document.
These steps must be followed when installing a replacement IDC Computer:
1. Disassembly (removal) of the existing IDC Computer.
2. Reassembly (installation) of the replacement IDC Computer.
3. Reload IDC software from the IDC Application Software CD-ROM used for the current installation.
4. Restore the detector calibration files by applying a "Restore IDC", see Job Card ELE A052 - Back-up/
Restore Procedure on page 771.
5. Select Service desktop/Utilities and perform the Detector Replacement procedure.
6. Reset the IDC Computer by powering it OFF/ON.

Table of Contents

Other manuals for GE Senographe DS

Related product manuals