GE Healthcare Senographe DS
Revision 1 Service Information and Procedures Class A 2385072-16-8EN
Job Card D/R A321 - IDC Computer
Page no. 1507 Chapter 9
JC-DR-A-321_DS.fm
6-3 Disassembly of IDC Computer (Dione4 version) on V3 Control Stations
1. Locate the IDC Computer (1) and UPS unit (2) inside the Control Station.
2. Make a note of all the cables connected to the UPS unit rear panel.
3. Disconnect all the power cables and data cable from the UPS unit rear panel.
4. Disconnect Ground cable (cross-head screwdriver) from the UPS unit rear panel.
5. Open the velcro straps (3) securing the UPS to the Control Station.
6. Remove the UPS unit from the Control Station by lifting it vertically from the framework and pulling it
towards you.
1
2
3