GE Healthcare Senographe DS
Revision 1 Service Information and Procedures Class A 2385072-16-8EN
Job Card D/R A321 - IDC Computer
Page no. 1515 Chapter 9
JC-DR-A-321_DS.fm
Reference Cable Label/Description Reference Cable Label/Description
1 IDC - Gantry Serial Link RT (1)
2240227 or 2244615
5 To Ethernet Switch:
2389288-2 on Penduick and above systems
or
2389288 on Pre-Penduick systems
2 IDC - Gantry Conditioner Link
2226612
6 AC mains
3 Detector Power Supply 2226621 7 Optical Fiber cables (to detector)
4 IDC - Gantry Serial Link RT
2240227 or 2246615
8 Tray ground wire
Dione4
2
3
6
7
54
1
2
3
1
6
7
4
Dione5
Note: For Dione5, an empty Ethernet port exists to the left hand side of the USB ports. The Ethernet cable from the switch must be connected
to the lowest Ethernet port, which is to the left hand side of the IDC - Gantry Conditioner Link.