GE Healthcare Senographe DS
Revision 1 Service Information and Procedures Class A 2385072-16-8EN
Job Card D/R A411 - Mains Distribution Board 200-PL1
Page no. 1605 Chapter 9
JC-DR-A-411.fm
6-2 Installation of the Mains Distribution Board 200-PL1
1. Carefully place the new Mains Distribution board on the Generator chassis. Ensure that the holes for
the four bolts line up with the four threads on the Generator chassis.
2. Attach the four bolts (1) (7 mm allen wrench) to secure the Mains Distribution board to the Generator
chassis.
3. Connect all the cables to the new Mains Distribution board as summarized in the diagram and table
below.
Connec-
tor Label
Tool Description of cable
(usually identified by code on a white label)
Cable
color
XJ1 8 mm open ended wrench XJ1 Brown
XJ2 8 mm open ended wrench Two cables: XJ2 Brown
XJ3 8 mm open ended wrench XJ3 Brown
XJ4 8 mm open ended wrench XJ4 Brown
XJ6 8 mm open ended wrench XJ6 Brown
XJ7 8 mm open ended wrench XJ7 Brown
XJ8 8 mm open ended wrench XJ8 Brown
XJ9 8 mm open ended wrench XJ9 Brown
XJ10 N/A XJ10/PL1 2111145 White
XJ11 N/A 200 PL1 XJ1 Gray
XJ12 N/A Empty N/A
XJ13 N/A Empty N/A
XJ14 N/A Empty N/A
TR1 - 5 8 mm open ended wrench SEL Brown
TR1 - 0 8 mm open ended wrench 0V Brown
1

Table of Contents

Other manuals for GE Senographe DS

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Senographe DS and is the answer not in the manual?

GE Senographe DS Specifications

General IconGeneral
BrandGE
ModelSenographe DS
CategoryMedical Equipment
LanguageEnglish

Related product manuals