Job Card D/R A411 - Mains Distribution Board 200-PL1
Page no. 1607 Chapter 9
JC-DR-A-411.fm
rotary switch in the Generator cabinet (position I).
5. Switch on the Senographe system electrical power (i.e. Generator, Gantry, and Control Station).
Wait until the Gantry boot is complete, and check that no error has been reported. If the Mains Distri-
bution board is operating correctly, the LED states on the Mains Distribution board are as follows:
red DE1 LED is OFF
green DE2 LED is ON
green DE3 LED is ON
red DE4 LED is OFF
yellow DE5 LED is OFF
yellow DE6 LED is ON
red DE7 LED is OFF
clear N1 LED is ON
clear K1 LED is ON
6. Reinstall the Generator covers (see Job Card PHY A042 - Remove/Reinstall Generator Covers on
page 513).
7 COMPLETION