GE Healthcare Senographe DS
Revision 1 Service Information and Procedures Class A 2385072-16-8EN
Job Card D/R A421 - Conditioner Unit
Page no. 1637 Chapter 9
JC-DR-A-421.fm
• Check that the Conditioner display shows the message IDC CONTROL for about 5 seconds shortly
after the system is switched on.
• Check that within about 15 minutes of operation manifold temperature shown on the Conditioner dis-
play is 26°C +
2°C.
• About 15 minutes after the system is operational, make an exposure; verify that the exposure pro-
ceeds correctly, with no warning message displayed after the acquisition.
6-6 Surplus Coolant Disposal
• In a normal non-sensitive environment, used and surplus coolant may be disposed of by pouring
down a drain.
• Triple-rinse containers before placing them in appropriate waste receptacles.
7 COMPLETION
None