EasyManua.ls Logo

GE Senographe DS - Page 1658

GE Senographe DS
1810 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Chapter 9 Page no. 1658
JC-CAL-A-003.fm
GE Healthcare Senographe DS
Revision 1 Service Information and Procedures Class A 2385072-16-8EN
Job Card CAL A003 - X-Ray Tube Focal Bias Voltage Scale Factor
a. Select SETUP/SETUP/3rd pt/CALIB on the X-ray Console; press the Exposure button on the
X-ray Console (no X-rays are produced) and hold it until the voltmeter reading stabilizes.
b. Note the focal bias voltage value measured from the voltmeter; release the Exposure button .
c. Select SETUP/MEASV on the X-ray Console.
d. Enter and confirm the measured focal bias voltage value.
7. Fourth calibration point:
a. Select SETUP/SETUP/4th pt/CALIB
on the X-ray Console; press the Exposure button on the
X-ray Console (no X-rays are produced) and hold it until the voltmeter reading stabilizes.
b. Note the focal bias voltage value measured from the voltmeter; release the Exposure button .
c. Select SETUP/MEASV on the X-ray Console.
d. Enter and confirm the measured focal bias voltage value.
8. Select SETUP/SETUP
/CALCUL/VALID on the X-ray Console. This calculates the parameters K, L,
M and N. The message CALCUL DONE appears on the X-ray Console display.
9. Press SETUP/SETUP/SETUP/SETUP/SETUP/SETUP/CKSUM/CKSUM on the X-ray Console to
perform a checksum.
6-4 Calibration Accuracy Check
1. Shut down the Senographe system from the Browser.
2. Re-connect flat cable WG8 to the Generator Command Board 400-PL1 connector XJ7.
3. Power on the Senographe system from the X-ray Console.
4. First check point:
a. Select SETUP/GENE/INSTAL/GENE/GENE/BIAS/VOLTAGE/CHANGE
on the X-ray Console
and note the actual value associated to the tube (see step 3 section 6-2);
b. Enter 3.000E+1V and valid.
c. Press SETUP/SETUP/SETUP/SETUP/CKSUM/CKSUM on the X-ray Console to perform a
checksum.
d. Press SETUP/SETUP/SETUP/SETUP (i.e. 4 times) on the X-ray Console.
e. Select 30kV/LARGE/MO/MO on the X-Ray Console.
f. Do a preparation with the Prep button on the X-ray Console.
g. Check that the result is between 28 and 32Vdc.
5. Second check point:
a. Select
SETUP/GENE/INSTAL/GENE/GENE/BIAS/VOLTAGE/CHANGE on the X-ray Con-
sole.
b. Enter 7.000E+1V and valid.
c. Press SETUP/SETUP/SETUP/SETUP/CKSUM/CKSUM on the X-ray Console to perform a
checksum.
d. Press SETUP/SETUP/SETUP/SETUP (i.e. 4 times) on the X-ray Console.
e. Select 30kV/LARGE/MO/MO on the X-ray Console.
f. Do a preparation with the Prep button on the X-ray Console.
g. Check that the result is between 67 and 73Vdc.
6. Third check point:

Table of Contents

Other manuals for GE Senographe DS

Related product manuals