Chapter 9 Page no. 1664
JC-CAL-A-009.fm
GE Healthcare Senographe DS
Revision 1 Service Information and Procedures Class A 2385072-16-8EN
Job Card CAL A009 - X-Ray Tube mA Measurement
6-1 Preparation
1. Turn off the Senographe.
2. Remove strap between terminal lugs MES_M_NEG and MES_M_POS of generator kV/mA board
701-PL1.
3. Connect the milliammeter between test points MES_M_NEG and MES_M_POS of generator kV/mA
board 701-PL1. Set the milliammeter to its 100 mA DC scale.
4. Turn on the Senographe.
Strap
MES_M_NEG
MES_M_POS