EasyManuals Logo

GE Senographe DS User Manual

GE Senographe DS
1810 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #1682 background imageLoading...
Page #1682 background image
Chapter 9 Page no. 1682
JC-CAL-A-027.fm
GE Healthcare Senographe DS
Revision 1 Service Information and Procedures Class A 2385072-16-8EN
Job Card CAL A027 - Tube Tilt Calibration
6-2 Procedure
Note:
If for any reason it is necessary to power OFF during the calibration procedure, you must power
OFF the whole system by using the Power OFF button on the X-ray Console.
1. From the Browser, launch the Service Desktop (see Launching the Service Desktop on page 120).
2. From the Service Desktop, click the Calibration button .
3. Select System/Tube Tilt.
4. Remove the Bucky, the compression paddle and any other objects from the field of view.
5. Install the Magstand simulator plug.
6. Click on START; make an exposure.
7. Wait for the result of the calibration. The calculated real tilt angle appears, together with the chest
wall to anode heel distance (for information only).
The real tilt angle must be 13.5°±0.5°.
8. Take the appropriate action:
- If the result is within the specified tolerance, click SAVE, The calibration is complete
- If there is an error message, click Cancel, then follow the instructions on the screen and repeat
the calibration.
When the calibration is complete, the AWS screen automatically returns to the System Calibration menu.
7 COMPLETION
If you are handing the machine back to the user at this point, power off the whole system from the X-ray
Console, then on again.

Table of Contents

Other manuals for GE Senographe DS

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Senographe DS and is the answer not in the manual?

GE Senographe DS Specifications

General IconGeneral
BrandGE
ModelSenographe DS
CategoryMedical Equipment
LanguageEnglish

Related product manuals