GE Healthcare Senographe DS
Revision 1 Service Information and Procedures Class A 2385072-16-8EN
Job Card CAL A041 - Bad Pixel Calibration
Page no. 1695 Chapter 9
JC-CAL-A-041.fm
tion specific to the detector.
For more information about Bad Pixel troubleshooting, refer to Job Card TSG A010 - Troubleshooting
Bad Pixel Calibration Failure on page 1019.
6-4 Gain Calibration
After completing the Bad Pixel calibration, Gain Calibration must be performed for all techniques (per-
form Job Card CAL A043 - Detector Gain Calibration on page 1697).
7 COMPLETION
If you are handing the machine back to the user at this point, switch the Senographe off, then on again.