GE Healthcare Senographe DS
Revision 1 Service Information and Procedures Class A 2385072-16-8EN
Job Card CAL A053 - mAs Non-linearity Calibration
9. Exit the Calibration Menu (SETUP/SETUP/SETUP//SETUP/SETUP).
6-3 Acceptable Values
• The value displayed for MASPRE CORR should be within the limits 1.00 and 1.20.
• If it is outside this range, check the Generator calibration and repeat the calibration.
7 COMPLETION
If you are handing the machine back to the user at this point, switch the Senographe off, then on again.