Chapter 11 Page no. 1766
SC-PM-A-001.fm
GE Healthcare Senographe DS
Revision 1 Service Information and Procedures Class A 2385072-16-8EN
Scenario PM A001 - Planned Maintenance Steering Guide
2 ACTIONS PRIOR TO PM VISIT
• Check that any applicable mandatory FMIs have been carried out.
Refer to Product history and compatibility on page 51.The following list may not be complete:
• Carry out the following tasks in preparation for the visit:
FMI Required Applied
FMI 12083 - Reliability for FB images in Stereotaxy
FMI 12075 - Safety : AWS 43.10/53.10
Task Title Task Definition
Obtain site status information and error logs Job Card PM A101 - Site Status: p. 1773
Analyze AWS Error Codes ERR SUB A005 - AWS Error Messages: p. 935
Analyze IDC Error Codes ERR SUB A007 - IDC Error Messages: p. 937
Analyze Gantry Error Codes ERR SUB A003 - Gantry Error Messages: p. 851
Analyze Generator Error Codes ERR SUB A001 - Generator Error Messages: p. 831